Calibration Certificate Options
JJ Calibrations offers flexible calibration options to meet
your exact requirements.
In order for us to properly service your instrumentation, we
must have a current Calibration Level Confirmation on file.
Standard Calibration: Instrumentation is calibrated to manufacture specifications with NIST traceable standards
in full compliance of ANSI Z540-1. Error direction and magnitude is reported if instrument is found out
of tolerance.
With Data: Instrumentation is calibrated to manufacture specifications with NIST traceable standards
in full compliance of ANSI Z540-1. A record of measurements is recorded and shown on the calibration
certificate. If the instrumentation is "in-tolerance", then no adjustments are made.
17025 Calibration
Service: Instrument is calibrated within our approved
scope of accreditation including data reporting and associated uncertainties of measurement
reporting.
Certificate Options
|
|
Without
Data
|
With
Data
|
Accredited 17025
With
Uncertainties
|
ANSI Z540-1
Compliant
|
X
|
X
|
X
|
ISO 9000:2001
Certification
|
X
|
X
|
X
|
ISO/IEC 17025 Compliant
Calibration
|
X
|
X
|
X
|
ISO/IEC 17025 Accredited
Calibration
|
|
|
X
|
NIST Traceable
Calibration
|
X
|
X
|
X
|
Min. 4:1 Test Uncertainty Ratio
(TUR)¹
|
X
|
X
|
X
|
Out of Tolerance (OOT)
Notification
|
X
|
X
|
X
|
Test Data
|
OOT²
|
X
|
X
|
Adjustments to within Manufacture
Specs³
|
X
|
X
|
X
|
Online Electronic Certificate
(DATA)
|
X
|
X
|
X
|
Test Point Specific Uncertainty
Data
|
|
|
X
|
Notes:
1- Exceptions of 4:1 TUR will be stated at the time we quote
service
2- If found OOT, magnitude and
direction of error reported
3- Adjustments are not made unless
found OOT. No charge for adjustments when made by JJ Calibrations
|
|